LAS OEM Sensors
Our technology - Your brand
We don’t have to put our name on it!
Instead we offer OEM solutions with our top-notch gas analysis technology for your product. We are happy to compliment your product solution with our know-how to create the best value for your customer. Wether we are extending your product functions or making it safer with e.g. continuous emission monitoring. We look forward to working with you!
long-term stability, high accuracy
LAS OEM-Sensors
LAS-OEM Modules belong to the most accurate measuring methods for trace gas analysis for all industrial and process gases (e.g. VHP/vapourized hydrogen preoxide) into the ppt range. The determination of different gases in one measuring process is possible. Due to long-range cells and focused laser beams, only a small sample volume is required.
Examples of customer-specific projects using LAS technology
Here you can have a look at our previously developed analyzers in the LAS technology area. Please do not hesitate to contact us, even if your application is not listed. We develop individually based on your requirements!
LAS Sensor
LASmini Sensor
Technical Data
Technical data | LAS | LASmini hot |
---|---|---|
Ambient temperature | 10 - 35 °C (non condensing) | |
Gas temperature | up to 35 °C | up to 190 °C |
Gas flow | < 2,0 l/min(1) | |
Communication | CAN + Modbus TCP/IP + analog output 4 - 20 mA (optional) | |
Dimensions (L x W x H) | 543 x 236 x 185 mm | 372 x 255 x 160 mm |
Weight | 14 kg | 6 kg |
Power supply | 24 V DC | |
1 depends on required pressure and pump |
Specifications
Specifications | LAS | LASmini hot |
---|---|---|
Gas (other gases on demand) | H2O2 | NH3 |
Measuring range (FS)1 | 0 - 1.000 ppm | 0 - 1.000 ppm |
Limit of detection (LOD)2 | < 50 ppb (3σ) t90-t10 ≤ 30 s @ 1.000 ppm | < 0,1 ppm (2σ@t10-t90 ≤ 2 s) |
Linearity | ≤ ± 0,5 ppm or ± 1 % of measured value | ≤ ± 1 ppm oder ± 1 % of measured value |
Zero drift | ≤ 50 ppb in 8 h | ≤ ± 0,15 ppm in 8 h |
Span drift | ≤ ± 1 % FS in 8 h | ≤ ± 0,5 % FS in 8 h |
1 increased ranges on demand | ||
2 optional < 5 ppb (2σ) |